COMPARISON OF TWO 1 kN FORCE TRANSDUCERS AT CONTINUOUS CALIBRATION |
|---|
| M. Hiti |
- Abstract:
- This paper describes a comparison of simultaneous continuous force introduction to two strain-gauge force transducer in a force calibration machine. The continuous force was generated by a single loading/unloading step of the full load of the 1 kN deadweight force calibration machine. Loading rates from 0.3 kN/s to 3.8 kN/s were applied. The results are presented for different loading speeds and for different amplifier types (225 Hz and 4.8 kHz carrier frequency).
- Keywords:
- calibration; continuous; force; error
- Download:
- IMEKO-TC3-2022-053.pdf
- DOI:
- 10.21014/tc3-2022.053
- Event details
- IMEKO TC:
- TC3
- Event name:
- TC3 Conference 2022
- Title:
24th Conference on the Measurement of Force, Mass and Torque (together with the 14th TC5 Conference on the Measurement of Hardness, the 6th TC16 Conference on Pressure and Vacuum Measurement, and the 5th TC22 Conference on Vibration Measurement)
- Place:
- Cavtat-Dubrovnik, CROATIA
- Time:
- 11 October 2022 - 13 October 2022