Implementation of BIPM Calibration Values to Calibration of A Solid State DC Voltage Standard in Laboratory of National Measurement Standard for Electricity and Time – BSN (SNSU – BSN)

Mohamad Syahadi, Hayati Amalia, Agah Faisal
Abstract:
A method for calibrating a solid state DC voltage standard in Laboratory of National Measurement Standards for Electricity and Time (NMS ET Lab) of National Standardization Agency of Indonesia is presented in this paper. The measurement and its evaluation are carried out based on differential measurement method to the test of solid state DC voltage standard by implementing calibration values from Bureau International des Poids et Mesures (BIPM) at the nominals of 1.018 V and 10 V. Corrections due to temperature, pressure, humidity coeficient and drift are also examined. Based on evaluation result, actual values for the 1.018 V and 10 V of the test were found to be 1.0179897 V and 9.999857 V with their respective measurement uncertainties of 1 μV and 17 μV.
Keywords:
Actual value, Calibration, Differential measurement method, Solid state DC voltage standard, Measurement uncertainty
Download:
IMEKO-TC10-2020-021.pdf
DOI:
-
Event details
IMEKO TC:
TC10
Event name:
TC10 Conference 2020 (ONLINE)
Title:

17th IMEKO TC10 Conference "Global trends in Testing, Diagnostics & Inspection for 2030” (2nd Conference jointly organized by IMEKO and EUROLAB aisbl)

Place:
Dubrovnik, CROATIA
Time:
20 October 2020 - 22 October 2020