EXPERIMENTAL VALIDATION OF A GENERAL-PURPOSE NON-LINEAR DYNAMIC MODEL FOR S/H-ADC DEVICES |
|---|
| P. A. Traverso, G. Pasini, D. Mirri, F. Filicori, G. Iuculano |
- Abstract:
- Experimental results are provided in this paper in order to validate the Discrete-Time Convolution approach, recently proposed for the modelling of both static and dynamic non-linearities in S/H-ADC devices. The comparison between actual device response and model predictions is provided, showing a very good agreement under operation conditions which are not suitably described by a conventional model, based on the characterisation of the static non-linearity only. A brief discussion about the possibility of embedding into the model almost any recent empirical technique for the characterisation of static sources of uncertainty within the device is proposed as well.
- Keywords:
- Analogue-to-Digital Converter, Sample-Hold, Non-Linear Dynamic Model, Volterra Series
- Download:
- IMEKO-IWADC-2003-57.pdf
- DOI:
- -