HIGH RESISTANCE COMPARISON AMONG KRISS, VNIIM AND NIM

K. M. Yu, J. C. Ryu, Y. P. Semenov, B. Litvinov, S. Haiming
Abstract:
In a high resistance comparison among KRISS, VNIIM and NIM, the relative resistance values of measurement results agreed well within 2 ppm in 10 megohm and 5 ppm in 1 gigohm. Also, relative combined standard uncertainties of the three institutes for the values is not beyond 4 ppm in 10 megohm and 8 ppm in 1 gigohm.
Keywords:
high resistance comparison, 10 megohm and 1 gigohm, adapted Wheatstone bridge, binary voltage divider bridge
Download:
IMEKO-WC-2000-TC8-P230.pdf
DOI:
-
Event details
Event name:
XVI IMEKO World Congress
Title:

Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

Place:
Vienna, AUSTRIA
Time:
25 September 2000 - 28 September 2000