1 MHZ MODULE TRANSFORMER BRIDGE

M. D. Klionskiy
Abstract:
A transformer bridge to study low-loss dielectrics at the frequency 1 MHz is developed. The main components of the bridge: a ratio transformer and variable measuring standards of capacitance and conductance are made in the form of removable modules. The bridge measurement range by loss angle tangent D is 10-6 - 10-3 with the capacitance 1-1000 pF. Measurement uncertainty D is ± (2-3) &itimes; 10-6, k = 2.
Keywords:
transformer bridge, loss angle
Download:
IMEKO-WC-2000-TC8-P222.pdf
DOI:
-
Event details
Event name:
XVI IMEKO World Congress
Title:

Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

Place:
Vienna, AUSTRIA
Time:
25 September 2000 - 28 September 2000