VECTOR DC MAGNETOMETERS CALIBRATION ERROR |
|---|
| V. Korepanov, F. Dudkin, K. Pajunpää |
- Abstract:
- The calibration system for vector DC magnetometers of Nurmijarvi Geophysical Observatory (Finland) is described. It is based on a three axes reference coil system and Observatory reference magnetometer. It is shown experimentally that the system allows to determine calibrated magnetometer parameters with an error about 0.02%. One important error source -influence of noncollinearity of reference and calibrated magnetometer axes - is analyzed and value of this error is estimated. Also additional possibilities of the Observatory infrastructure are described: determination of temperature errors of magnetometer’s sensor and electronics and long-term temporal drift study of up to 9 magnetometers simultaneously.
- Keywords:
- magnetometer, calibration, axes collinearity, error
- Download:
- IMEKO-WC-2000-TC4-P137.pdf
- DOI:
- -
- Event details
- Event name:
- XVI IMEKO World Congress
- Title:
Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future
- Place:
- Vienna, AUSTRIA
- Time:
- 25 September 2000 - 28 September 2000