THE EFFECT OF HIGH TRANSVERSE INPUTS ON ACCELEROMETER CALIBRATION |
|---|
| Richard W. Bono, Eric J. Seller, Mark I. Schiefer |
- Abstract:
- ISO 16063 part 21 defines the back-to-back comparison technique for accelerometer calibration. Included in its most recent revision is a recommendation for acceptable limits on shaker transverse motion characteristics. The effect of high transverse inputs can be devastating to accurate accelerometer calibration. This paper discusses the differences between mechanical flexurebased electrodynamic shakers and air bearing shakers and the resulting effects on calibration accuracy and uncertainty.
- Keywords:
- calibration, transverse, accelerometer