A PRECISION CALIBRATION SET-UP FOR AC MAGNETIC FLUX DENSITY MEASUREMENT IN THE RANGE OF 1 Hz TO 20 kHz |
|---|
| Po Gyu Park, Young Gyun Kim, Wan-Seop Kim, V. N. Kalabin, Vladlen Ya. Shifirn |
- Abstract:
- A precision generating and measurement system of AC magnetic flux density was developed by using a linear low-frequency AC magnetic field converter, a quartz Helmholtz coil, a voltage comparator, a multimeter, a standard resistor, and a computer with a program for automatic measurement. The expanded uncertainty (k = 2) of the calibration system is less than 0.2 % over the frequency range of 1 Hz to 20 kHz.
- Keywords:
- AC magnetic flux density, Calibration
- Download:
- IMEKO-WC-2009-TC4-197.pdf
- DOI:
- -
- Event details
- Event name:
- XIX IMEKO World Congress
- Title:
Fundamental and Applied Metrology
- Place:
- Lisbon, PORTUGAL
- Time:
- 06 September 2009 - 11 September 2009