DEVELOPMENT OF AN AUTOMATIC CALIBRATION SYSTEM FOR CLINICAL ELECTRICAL THERMOMETERS |
|---|
| Le Chen, Jian Sun, Yaqiong Fu, Hongwei Xu |
- Abstract:
- An automatic test system for clinical electrical thermometers was developed to enhance the efficiency of error tests. Besides the temperature measurement unit, mechanical drive and machine vision units were designed to implement automatic tests. An uncertainty analysis was also presented. According to the special measurement action of clinical thermometers, the Monte Carlo method was used to evaluate the uncertainty. The accuracy of recognition of indications was greater than 99% in experiments.
- Keywords:
- calibration system, clinical electrical thermometer, CCD
- Download:
- IMEKO-WC-2009-TC12-087.pdf
- DOI:
- -
- Event details
- Event name:
- XIX IMEKO World Congress
- Title:
Fundamental and Applied Metrology
- Place:
- Lisbon, PORTUGAL
- Time: