FULLY DIGITAL FLUXGATE MAGNETOMETER – FIRST RESULTS

A. Cerman, A. Kuna, P. Ripka
Abstract:
Presented paper describes design, implementation and first results of the fully digital fluxgate magnetometer. The magnetometer operates in feedback configuration and uses fully digital signal detection realized in the digital signal processor (DSP). The magnetometer has measuring range ±100 μT, the linearity error is less than 40 ppm of FSR, noise 1.08 nTRMS/√Hz@1Hz and the frequency range of measured magnetic field is 15 Hz.
Download:
IMEKO-TC4-2004-031.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
TC4 Symposium 2004
Title:
XIII IMEKO TC4 International Symposium on Measurements for Research and Industrial Applications (together with IXth International Workshop on ADC Modeling and Testing, IWADC)
Place:
Athens, GREECE
Time:
29 September 2004 - 01 October 2004