CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO 16063-41

Uwe Buehn, Holger Nicklich, Heinz Weissing, Georg Siegmund
Abstract:
All users of primary calibration systems - excepting the national metrological institutes - are under obligation to trace back their reference standards – their laser vibrometer standards – to their respective national standard if they intend to apply for (or maintain) accreditation or to strive for a very low measuring uncertainty for other reasons. A calibration method will be described in this paper that has been used by the accredited laboratory of SPEKTRA for calibrating a greater number of laser vibrometer standards. All these calibrations proved the reliability and the very low measurement uncertainty of this method.
Keywords:
calibration, laser vibrometer standard, primary